Specifications: |
Analog |
|
Performance: |
Economical
 |
Conventional
 |
Digital
 |
Linearity |
± 3.0% of Full Scale |
± 1.0% of Full Scale |
± 0.5% of Full Scale |
Repeatability & Hysteresis
Temperature Coefficient
Response Time
(to within 1% of set point)
Long Term Stability
(Continuous Operation) |
± 0.15% FS
0.09% / ºC from 21ºC set point
100 msec
0.2% FS / >7500 hours |
± 0.15% FS
0.009% / ºC from 21ºC set point
100 msec
0.2% FS / >7500 hours |
± 0.15% FS
0.09% / ºC Uncompensated
0.005%/ ºC Compensated
250 msec
0.2% FS / >7500 hours |
Fluid Requirements: |
|
Calibration Gas
Flow Ranges
Pressure Range
Cal Pressure
Internal Leakage
External Leakage |
Customer Specified, Non Corrosive
Customer Specified, 0 sccm at Vc < 1% FS, Min sccm at Vc = 2% FS, Max sccm at Vc = 100% FS
0-100 Psig
25 Psi - Typical
10-5 sccm @ 1, 30 and 100 Psi (He)
2 x 10-6 sccm (He) |
Electrical Requirements: |
|
Input Voltage (single supply)
Power Consumption
Control Voltage
Output Voltage |
12, 15, 18, or 24 volts +10% - 3%
2.0 Watts Maximum
0 - 5 VDC, 1 - 5 VDC, or 4-20mA
0 - 5 VDC |
Environmental Requirements: |
|
Operating Ambient Temp.
Storage Temperature
Humidity Range
Nominal Temperature |
0-50ºC
-40 - 70ºC
5% - 95% (Relative, non-condensing)
21ºC ±10ºC |
Mechanical Configuration: |
|
Input/Output Ports
Dimensions
Weight |
10-32 Threaded (Other sizes, consult factory)
L: 2.8 W: 1.5 H: 5.6
13.0 oz |
Construction (wetted materials): |
|
Manifold
Filter Screen
O-Rings
Tubing
Sealant
Flow Sensor |
Nickel Plated Aluminum, Brass 360
Polyester
Fluorocarbon (Viton®), others at additional cost
Polyurethane
Anaerobic Sealant
Engineering Thermoplastic, Silicon, Silicon Nitrite, Gold, Aluminum Oxide, Epoxy Sealant, Fluorocarbon (Viton®), Polyester |
Pneucleus Technologies LLC, 169 Depot Rd., Hollis, NH 03049, USA
Ph: (603) 465-7346 Fx: (603) 465-3594 info@pneucleus.com www.pneucleus.com
© Copyright 2006 Pneucleus Technologies LLC |